NPL builds long range alpha detector

The UK’s National Physical Laboratory (NPL) has developed a new portable radiation detector that can assess the safety of potentially contaminated areas far quicker than current methods. The prototype was inspired by the aftermath of the murder of Alexander Litvinenko in London 2006.

Litvinenko’s death was due to radiation poisoning from polonium-210. It prompted wide-scale work to assess radiation levels at almost 50 premises in London that he had been prior to his death before opening them up again to the public, costing the Health Protection Agency £2 million.

The clean up process took an extended period of time as scintillation counters were used to detect alpha particles and therefore measure the radiation levels. Alpha particles only have a range of about 2 cm and can be suppressed by surface moisture and roughness. Geiger counters also require a skilled operator to work, making it a painstaking and laborious process to scan wide areas very close to the surface.

NPL has developed a prototype device which can detect the optical photons generated by alpha particles in air, thus having a far greater range.

Alpha particles can cause ionization of nitrogen in air, which in turn produces fluorescence photons that show up as discrete spectral lines in the UV band of the electromagnetic spectrum. These photons can travel much further than alpha particles and are not hindered by surface particles or roughness and can be transmitted through any UV transparent material such as plastic or glass. This means an area could be made safe before a contamination assessment is carried out.

NPL has developed a portable device for low-level monitoring in normal lighting conditions. Laboratory tests have shown that the prototype is capable of detecting alpha induced nitrogen fluorescence at distances greater than the range of alpha particles (2 cm). The prototype has demonstrated detection up to 20 cm, limited currently by the optics on the device. However, the principle is completely scaleable with small devices having a range of cms and non-portable devices reaching kms.

Ray Chegwin, Business Development Manager at NPL, said:
“NPL has a strong heritage of using world class measurement science to benefit society and industry in the UK and beyond. By producing a prototype that can measure radiation up to ten times further away than current methods we can greatly improve the time it takes to safely assess any suspected radioactive areas. Our prototype could have saved the time it took to monitor the areas relevant to the Litvinenko case, and cut the costs of such an essential operation. Of course, the detector could be used to support radiation safety and decommissioning operations in the nuclear industry, amongst others, and we’re currently investigating these.”

Notes to editors

About the NPL Alpha Detector:

The spectral range of the fluorescence photons emitted due to alpha ionisation of nitrogen are located in the predominantly UV range (300 — 490 nm) of the optical spectrum. Tests have been carried out to eliminate any possibility of the detected signal coming from beta or gamma radiation.

The device can operate in a sodium lamp room. In this environment, the light levels are similar to normal room light levels (lights on). The filtered detector is entirely blind to the wavelength of light emitted by sodium lights. The detector is also capable of detecting the fluorescence signal through any UV transparent material.

The optical housing has been designed so that any 1-inch optics can be inserted to select particular wavelengths. The device is modular; adapters could be made for larger optics, larger area detectors can also be used. Custom electronics have been designed and implemented to make the device portable. A lithium ion battery provides up to 8 hours of usage in field-based applications.

The National Physical Laboratory

The National Physical Laboratory (NPL) in Teddington is one of the UK’s leading science facilities and research centres. It is a world-leading centre of excellence in developing and applying the most accurate standards, science and technology available.
NPL occupies a unique position as the UK’s National Measurement Institute and sits at the intersection between scientific discovery and real world application. Its expertise and original research have underpinned quality of life, innovation and competitiveness for UK citizens and business for more than a century:

  • NPL provides companies with access to world leading support and technical expertise, inspiring the absolute confidence required to realise competitive advantage from new materials, techniques and technologies
  • NPL expertise and services are crucial in a wide range of social applications – helping to save lives, protect the environment and enable citizens to feel safe and secure. Support in areas such as the development of advanced medical treatments and environmental monitoring helps secure a better quality of life for all
  • NPL develops and maintains the nation’s primary measurement standards, supporting an infrastructure of traceable measurement throughout the UK and the world, to ensure accuracy and consistency.

Substack subscription form sign up